踝关节生物力学实验放置压敏片入路新方法
摘要点击次数: 1748   全文下载次数: 1211   投稿时间:2005-11-17    
作者Author单位AddressE-Mail
韩长伶 HAN Chang-ling 河北医科大学第三医院骨科研究所,河北石家庄050051 Institute of Orthopaedic,the Third Hospital of Medical University of Hebei,Shijiazhuang 050051,Hebei,China  
姚双权 YAO Shuang-quan 河北医科大学第三医院骨科研究所,河北石家庄050051 Institute of Orthopaedic,the Third Hospital of Medical University of Hebei,Shijiazhuang 050051,Hebei,China  
刘勇 LIU Yong 河北医科大学第三医院骨科研究所,河北石家庄050051 Institute of Orthopaedic,the Third Hospital of Medical University of Hebei,Shijiazhuang 050051,Hebei,China  
张奉琪 ZHANG Feng-qi 河北医科大学第三医院骨科研究所,河北石家庄050051 Institute of Orthopaedic,the Third Hospital of Medical University of Hebei,Shijiazhuang 050051,Hebei,China  
期刊信息:《中国骨伤》2006年,第19卷,第4期,第223-225页
DOI:doi:10.3969/j.issn.1003-0034.yyyy.nn.zzz
基金项目:河北省科技攻关项目资助(编号03276196D-21)
中文摘要:

目的:探讨踝关节生物力学试验放置压敏片的新方法,为踝关节生物力学试验提供依据。

方法:采用6具防腐尸体标本,左足组为A组,从内踝放入压敏片,右足组为B组,从外踝放入压敏片。标本固定在实验机上,轴向压缩700N,维持120秒,分析压敏片。

结果:A、B组踝关节接触面积分别是(284.8±27.3)、(281.3±24.7)mm2;踝关节前内、前外、后内、后外受力面积A组分别是(77.4±8.3)、(67.8±6.2)、(78.1±6.5)、(61.5±9.2)mm2,B组分别是(81.5±9.2)、(64.2±5.9)、(80.9±6.1)、(54.7±8.2)mm2;平均压强A组分别是(2.09±0.06)、(2.30±0.48)、(1.73±0.41)、(1.87±0.43)MPa,B组分别是(2.11±0.47)、(2.33±0.45)、(1.70±0.38)、(1.82±0.40)MPa;峰值压强A组分别是(3.24±0.44)、(3.38±0.47)、(2.89±0.38)、(3.02±0.41)MPa,B组分别是(3.29±0.43)、(3.36±0.49)、(2.85±0.34)、(3.05±0.44)MPa。A、B组比较差异均无统计学意义(P>0.05)。

结论:经内踝入路放置压敏片试验效果同经外踝入路放置压敏片相当,为踝关节力学研究提供了一个新方法。
【关键词】踝关节  压敏片  生物力学
 
A new method to place pressure sensitive film on ankle joint:An experiment of biomechanics
ABSTRACT  

Objective:To discuss a new method to place pressure sensitive film on ankle joint through biomechanical experiment.

Methods:The six specimens of anticorrosive cadaver were divided into two groups(A and B)according to left and right foot.The pressure sensitive film were respectively placed into medial malleolus(group A)and lateral malleolus(group B).They were tested on experimental machine,the force of axio-compression was 700 N,keeping 120 seconds.

Results:The contact surface of ankle joint were(284.8±27.3)mm2,(281.3±24.7)mm2 in group A and group B,respectively.In group A,the compressive surface of ankle joint in intra-anterior,extra-anterior,intra-lateral,extra-lateral were(77.4±8.3)mm2,(67.8±6.2)mm2,(78.1±6.5)mm2,(61.5±9.2)mm2,respectively;and in group B,they were(81.5±9.2)mm2,(64.2±5.9)mm2,(80.9±6.1)mm2,(54.7±8.2)mm2,respectively.The average pressure were(2.09±0.06)MPa,(2.30±0.48)MPa,(1.73±0.41)MPa,(1.87±0.43)MPa in group A,respectively;and(2.11±0.47)MPa,(2.33±0.45)MPa,(1.70±0.38)MPa,(1.82±0.40)MPa in group B,respectively.The peak pressure were(3.24±0.44)MPa,(3.38±0.47)MPa,(2.89±0.38)MPa,(3.02±0.41)MPa in group A,respectively;and(3.29±0.43)MPa,(3.36±0.49)MPa,(2.85±0.34)MPa,(3.05±0.44)MPa in group B(P>0.05),respectively.

Conclusion:Placing pressure sensitive film into medial malleolus can obtain the same results as those placing pressure sensitive film into lateral malleolus,thus provides a new method for biomechanical experiment of ankle joint.
KEY WORDS  Ankle joint  Pressure sensitive film  Biomechanics
 
引用本文,请按以下格式著录参考文献:
中文格式:韩长伶,姚双权,刘勇,张奉琪.踝关节生物力学实验放置压敏片入路新方法[J].中国骨伤,2006,19(4):223~225
英文格式:HAN Chang-ling,YAO Shuang-quan,LIU Yong,ZHANG Feng-qi.A new method to place pressure sensitive film on ankle joint:An experiment of biomechanics[J].zhongguo gu shang / China J Orthop Trauma ,2006,19(4):223~225