Sponsor
  • ·
  • Chinese Association of
    Integrative Medicine;
    China Academy of Chinese
    Medicine Sciences
Editing
  • ·
  • Editorial Board of
    China Journal of
    Orthopaedics and Traumatology
Publishing
  • ·
  • Publishing House,
    China Journal of
    Orthopaedics and Traumatology
Overseas Distributor
  • ·
  • China International Book
    Trading Corporation
    P.O.Box 399,Beijing,China
    Code No.M587
Mail-order
  • ·
  • Publishing House,
    China Journal of
    Orthopaedics and Traumatology
    No.16A, Nanxiaojie, Dongzhimennei,
    Beijing 100700,China
    Tel:0086-10-84020925
    Fax:0086-10-84036581
    Http://www.zggszz.com
    E-mail:zggszz@sina.com
股骨小转子骨折固定复位器的设计与临床应用
Hits: 2036   Download times: 1239   Received:October 20, 2014    
作者Author单位UnitE-Mail
叶锋 YE Feng 上虞市中医院骨伤科, 浙江 上虞 312300 Department of Orthopaedics, Shangyu Hospital of Traditional Chinese Medicine, Shangyu 312300, Zhejiang, China yf941310@163.com 
张龙君 ZHANG Long-jun 上虞市中医院骨伤科, 浙江 上虞 312300 Department of Orthopaedics, Shangyu Hospital of Traditional Chinese Medicine, Shangyu 312300, Zhejiang, China  
李英周 LI Ying-zhou 上虞市中医院骨伤科, 浙江 上虞 312300 Department of Orthopaedics, Shangyu Hospital of Traditional Chinese Medicine, Shangyu 312300, Zhejiang, China  
郑晓东 ZHENG Xiao-dong 上虞市中医院骨伤科, 浙江 上虞 312300 Department of Orthopaedics, Shangyu Hospital of Traditional Chinese Medicine, Shangyu 312300, Zhejiang, China  
王晓 WANG Xiao 上虞市中医院骨伤科, 浙江 上虞 312300 Department of Orthopaedics, Shangyu Hospital of Traditional Chinese Medicine, Shangyu 312300, Zhejiang, China  
杨永博 YANG Yong-bo 上虞市中医院骨伤科, 浙江 上虞 312300 Department of Orthopaedics, Shangyu Hospital of Traditional Chinese Medicine, Shangyu 312300, Zhejiang, China  
陈园升 CHEN Yuan-sheng 上虞市中医院骨伤科, 浙江 上虞 312300 Department of Orthopaedics, Shangyu Hospital of Traditional Chinese Medicine, Shangyu 312300, Zhejiang, China  
期刊信息:《中国骨伤》2015年28卷,第8期,第726-729页
DOI:10.3969/j.issn.1003-0034.2015.08.011


目的:探讨小转子复位固定在转子间骨折治疗中的临床意义。

方法:自2012年1月至2013年12月,治疗32例不稳定性转子间骨折患者,其中男11例,女21例;年龄45~81岁,平均64岁。EvansⅡ型12例,Ⅲ型13例,Ⅳ型7例。均采用自行设计的爪式可控复位固定器进行小转子复位。方法:单纯使用拉力螺钉19例,使用钢丝13例。通过观察患者髋内翻、股骨头颈切割、钉板断裂等情况,以及髋关节功能(采用Harris关节功能评分)来分析疗效。

结果:32例患者均获得随访,时间9个月~2年,平均13个月。32例患者中发生髋内翻1例,无股骨头颈切割及钉板断裂。髋关节功能评分为91.80±3.05.

结论:后内侧骨皮质连续性的重建和固定是影响股骨转子间骨折稳定性的关键因素,采用自行设计的爪式可控复位固定器可在术中有效的复位固定小转子,具有快速复位和固定、操作简捷的优点。
[关键词]:髋骨折  牵引术  骨折固定术,内
 
Design of the reduction fixator and its clinical application in treating fracture of lesser trochanter of femur
Abstract:

Objective:To discuss the clinical significance of the reduction and fixation for femoral lesser trochanteric fragment in treating the femoral comminuted intertrochanteric fractures.

Methods:From January 2012 to December 2013,32 patients with intertrochanteric fractures were treated by surgery,and self-designed reduction fixators were used in the patients for the fixation of lesser trochanter of femur. There were 11 males and 21 females,ranging in age from 45 to 81 years old with an average of 64 years old. According to the Evans typing,12 cases were typeⅡ,13 cases were type Ⅲ and 7 cases were type Ⅳ。 Simple lag screw fixed in 19 cases and steel wire fixed in 13 cases. Hip joint function was evaluated according Harris score and the complications such as coxa adducta,cutting of femoral head and neck,implants breakage were observed.

Results:Thirty-two patients were followed up from 9 to 24 months with an average of 13 months. Coxa adducta occurred in 1 case,no cutting of femoral head and neck occurred and implants breakage were found. The mean Harris score was 91.80±3.05 in 32 patients.

Conclusion:The reconstruction and fixation for the posterior medial bone cortex continuity is a key factor on affect the stability of femoral intertrochanteric fracture. Self-designed reduction fixators can effective reduce and fix the lesser trochanter of femur,it has advantage of fast reduction and fixation,and simple operation.
KEYWORDS:Hip fractures  Traction  Fracture fixation,interal
 
引用本文,请按以下格式著录参考文献:
中文格式:叶锋,张龙君,李英周,郑晓东,王晓,杨永博,陈园升.股骨小转子骨折固定复位器的设计与临床应用[J].中国骨伤,2015,28(8):726~729
英文格式:YE Feng,ZHANG Long-jun,LI Ying-zhou,ZHENG Xiao-dong,WANG Xiao,YANG Yong-bo,CHEN Yuan-sheng.Design of the reduction fixator and its clinical application in treating fracture of lesser trochanter of femur[J].zhongguo gu shang / China J Orthop Trauma ,2015,28(8):726~729
View Full Text  View/Add Comment  Download reader
Close




版权所有:Editorial Office of China Journal of Orthopaedics and Traumatology京ICP备12048066号  版权声明
地址:No.16A, Nanxiaojie, Dongzhimennei, Beijing 100700, China
电话:0086-10-84036581 传真:0086-10-84036581 Email:zggszz@sina.com