液晶数字显示角度仪在全髋关节置换术中的应用价值
摘要点击次数: 1609   全文下载次数: 493   投稿时间:2020-07-20    
作者Author单位AddressE-Mail
侯靖钊 HOU Jing-zhao 靖江市人民医院, 江苏 靖江 214500 Department of Orthopaedics, the People's Hospital of Jingjiang, Jingjiang 214500, Jiangsu, China  
王海红 WANG Hai-hong 靖江市人民医院, 江苏 靖江 214500 Department of Orthopaedics, the People's Hospital of Jingjiang, Jingjiang 214500, Jiangsu, China  
程彦骁 CHENG Yan-xiao 靖江市人民医院, 江苏 靖江 214500 Department of Orthopaedics, the People's Hospital of Jingjiang, Jingjiang 214500, Jiangsu, China  
包洪卫 BAO Hong-wei 靖江市人民医院, 江苏 靖江 214500 Department of Orthopaedics, the People's Hospital of Jingjiang, Jingjiang 214500, Jiangsu, China hjzhcy@163.com 
王禹基 WANG Yu-ji 常州市第二人民医院, 江苏 常州 213000  
孙钰 SUN Yu 苏北人民医院, 江苏 扬州 225001  
期刊信息:《中国骨伤》2021年,第34卷,第7期,第597-600页
DOI:10.12200/j.issn.1003-0034.2021.07.002
基金项目:江苏省"科教强卫"青年医学人才项目(编号:QNRC2016344)
中文摘要:

目的:探讨在全髋关节置换术中使用液晶数字显示角度仪控制前倾角的应用价值。

方法:回顾分析自2018年1月至2019年12月83例行初次全髋关节置换术的患者,其中男28例,女55例;年龄42~81(70.4±7.9)岁。股骨颈骨折63例,股骨头缺血性坏死20例。所有患者术中使用液晶数显角度仪控制髋臼杯假体的前倾角,术后采用CT扫描,测量髋臼杯的前倾角,两者进行比较,了解使用液晶数显角度仪的准确性。

结果:术后CT测量提示患者的髋臼前倾角均位于Lewinnek提倡的安全区内,术中使用液晶数显角度仪测量髋臼杯的前倾角度中位数为14.20°(12.80~15.40)°,术后CT扫描测量的髋臼杯的前倾角中位数为14.20°(13.40~15.50)°,两者比较差异无统计学意义(Z=-1.725,P=0.085)。

结论:应用液晶数显角仪器对术中控制髋臼杯的前倾角是一种准确可靠的方法,具有良好的辅助参考价值。
【关键词】关节成形术,置换,髋  前倾角  数显角度仪  直接前方入路
 
Application value of liquid crystal digital display goniometer in total hip arthroplasty
ABSTRACT  

Objective: To investigate the application value of liquid crystal digital display goniometer in total hip arthroplasty.

Methods: From January 2018 to December 2019,83 patients underwent primary total hip arthroplasty,including 28 males and 55 females,aged 42 to 81(70.4±7.9) years. There were 63 cases of femoral neck fracture and 20 cases of avascular necrosis of femoral head. All patients used liquid crystal digital goniometer to control the anteversion of acetabular cup prosthesis during operation,and CT scanning was used to measure the anteversion of acetabular cup after operation. The two methods were compared to understand the accuracy of using liquid crystal digital goniometer.

Results: Postoperative CT measurement showed that the acetabular anteversion of all patients was in the safe area advocated by Lewinnek. The anteversion angle of acetabular cup measured by liquid crystal digital goniometer was 14.20(12.80 to 15.40)°,and the anteversion angle of acetabular cup measured by postoperative CT scan was 14.20(13.40 to 15.50)°. There was no significant difference between the two(Z=-1.725,P=0.085).

Conclusion: It is an accurate and reliable method to control the anteversion of acetabular cup with liquid crystal digital display angle instrument,which has a good auxiliary reference value.
KEY WORDS  Arthroplasty,replacement,hip  Acetabular anteversion  Digital display goniometer  Direct anterior approach
 
引用本文,请按以下格式著录参考文献:
中文格式:侯靖钊,王海红,程彦骁,包洪卫,王禹基,孙钰.液晶数字显示角度仪在全髋关节置换术中的应用价值[J].中国骨伤,2021,34(7):597~600
英文格式:HOU Jing-zhao,WANG Hai-hong,CHENG Yan-xiao,BAO Hong-wei,WANG Yu-ji,SUN Yu.Application value of liquid crystal digital display goniometer in total hip arthroplasty[J].zhongguo gu shang / China J Orthop Trauma ,2021,34(7):597~600
阅读全文  下载  查看/发表评论  下载PDF阅读器
关闭




版权所有:《中国骨伤》杂志社京ICP备12048066号-2  版权声明
地址:北京市东直门内南小街甲16号,100700
电话:010-64089487 传真:010-64089792 Email:zggszz@sina.com

京公网安备 11010102004237号