小夹板固定系统的骨干轴向牵引作用分析
摘要点击次数: 2525   全文下载次数: 1669   投稿时间:2016-05-21    
作者Author单位AddressE-Mail
钟红刚 ZHONG Hong-gang 中国中医科学院望京医院, 北京 100700 Wangjing Hospital, China Academy of Chinese Medical Science, Beijing 100700, China zhglyw2013@163.com 
刘卫华 LIU Wei-hua 中国中医科学院望京医院, 北京 100700 Wangjing Hospital, China Academy of Chinese Medical Science, Beijing 100700, China  
董福慧 DONG Fu-hui 中国中医科学院望京医院, 北京 100700 Wangjing Hospital, China Academy of Chinese Medical Science, Beijing 100700, China  
期刊信息:《中国骨伤》2017年,第30卷,第1期,第57-59页
DOI:10.3969/j.issn.1003-0034.2017.01.013
基金项目:国家自然科学基金(编号:81573809);中国中医科学院自由探索项目(编号:zz0708093)
中文摘要:

目的:测量小夹板固定骨干骨折时骨干轴向牵引力。

方法:通过有机玻璃材料加工成类似哑铃形状模拟长骨两端膨大的特征,在模拟骨干骨折截断处串联安装力传感器;同时,采用布袋液压囊模拟肌肉等软组织,测量夹板固定时断端骨干轴向牵引力。

结果:轴向牵引力与横向夹板压力有正相关,产生的骨干轴向牵引力约为夹板横向加压力大小的1/10。

结论:筋膜包裹的肌肉与布袋液压囊力学作用类似,夹板在肢体侧面横向加压时,在骨折端可以产生骨干轴向牵引力。
【关键词】小夹板  骨折  
 
Analysis of the axial stretching force of the small splint fixation system
ABSTRACT  

Objective: To measure the axial stretching force produced by the small splint fixation system in the treatment of the fracture.

Methods: Dumbbell shaped plexiglass was machined to model the bone shaft with expanded two ends.Force transducer was set at the model fracture site to measure the stretching force produced by fracture fixation of the small splint with cloth aqueous bag that simulated the muscles and other soft tissue underneath.

Results: There was positive relationship between the axial stretching force produced at the model fracture site and the transverse pressing force exerted by the external plates.The ratio of which was 1/10.

Conclusion: Muscle enwrapped with fascia is similar to the cloth aqueous bag system in mechanical structure.From this article, axial stretching force exists at the bone shaft fracture site when fixed by small splint which exerts the force transversely from the outside of the extrimity.
KEY WORDS  Small splint  Fractures  Force
 
引用本文,请按以下格式著录参考文献:
中文格式:钟红刚,刘卫华,董福慧.小夹板固定系统的骨干轴向牵引作用分析[J].中国骨伤,2017,30(1):57~59
英文格式:ZHONG Hong-gang,LIU Wei-hua,DONG Fu-hui.Analysis of the axial stretching force of the small splint fixation system[J].zhongguo gu shang / China J Orthop Trauma ,2017,30(1):57~59
阅读全文  下载  查看/发表评论  下载PDF阅读器
关闭




版权所有:《中国骨伤》杂志社京ICP备12048066号-2  版权声明
地址:北京市东直门内南小街甲16号,100700
电话:010-64089487 传真:010-64089792 Email:zggszz@sina.com

京公网安备 11010102004237号