剪切波超声弹性成像技术在颈肩肌筋膜疼痛综合征中的应用
摘要点击次数: 2541   全文下载次数: 1280   投稿时间:2015-09-18    
作者Author单位AddressE-Mail
郭玲 GUO Ling 中国中医科学院望京医院超声科, 北京 100102 Wangjing Hospital, China Academy of Chinese Medical Sciences, Beijing 100102, China  
张晨 ZHANG Chen 中国中医科学院望京医院超声科, 北京 100102 Wangjing Hospital, China Academy of Chinese Medical Sciences, Beijing 100102, China lvsedehaiyang@163.com 
张丁丁 ZHANG Ding-ding 中国中医科学院望京医院超声科, 北京 100102 Wangjing Hospital, China Academy of Chinese Medical Sciences, Beijing 100102, China  
高景华 GAO Jing-hua 中国中医科学院望京医院超声科, 北京 100102 Wangjing Hospital, China Academy of Chinese Medical Sciences, Beijing 100102, China  
刘广会 LIU Guang-hui 中国中医科学院望京医院超声科, 北京 100102 Wangjing Hospital, China Academy of Chinese Medical Sciences, Beijing 100102, China  
王尚全 WANG Shang-quan 中国中医科学院望京医院超声科, 北京 100102 Wangjing Hospital, China Academy of Chinese Medical Sciences, Beijing 100102, China  
期刊信息:《中国骨伤》2016年,第29卷,第2期,第142-145页
DOI:10.3969/j.issn.1003-0034.2016.02.011
基金项目:
中文摘要:

目的:探讨剪切波超声弹性成像技术在颈肩肌筋膜疼痛综合征中的应用.

方法:选取2013年12月至2014年7月临床诊断为颈肩肌筋膜疼痛综合征的患者30例(病例组)及对照组30例(健康人群).病例组男17 例,女13例,平均年龄(44±3)岁;对照组男22 例,女8例,平均年龄(37±5)岁.病例组患者行手法治疗,隔日1次,共7次,分别以剪切波超声弹性成像技术检测病例组患者治疗前后斜方肌紧张处,以及对照组斜方肌降部肌腹,记录并比较组织弹性图和杨氏模量值,对结果进行统计分析.

结果:病例组治疗前斜方肌组织弹性图以绿蓝色为主,弹性评分3.70±1.53,杨氏模量值(43.4±15.6) kPa;治疗后组织弹性图以蓝色为主,弹性评分2.40±0.87,杨氏模量值(29.0±5.9) kPa;对照组组织弹性图以蓝色为主,弹性评分1.60±0.72,杨氏模量值(24.0±7.6) kPa;三者两两比较,P=0.000,差异均有统计学意义.

结论:剪切波超声弹性成像技术可作为检测中医手法治疗颈肩肌筋膜疼痛综合征疗效的客观检测方法,且较敏感.
【关键词】    肌筋膜疼痛综合征  手法,骨科  剪切波超声弹性成像
 
Application of shear wave elastography in the evaluation of neck-shoulder myofascial pain syndrome
ABSTRACT  

Objective:To study clinical value of shear wave elastography(SWE) in the evaluation of neck-shoulder myofascial pain syndrome.

Methods:From December 2013 to July 2014,30 patients diagnosed as neck-shoulder myofascial pain syndrome were in the treatment group,including 17 males and 13 females,with an average age of (44±3) years old. Thirty healthy people were in the control group,including 22 males and 8 females,with a mean age of (37±5) years old. The patients in the treatment group were treated with manipulation,once every other day,total 7 times. The SWE was used to detect tension part of trapezius muscle of patients in the treatment group before and after treatment,as well as to detect muscle belly at the descending part of trapezius muscle in the control group. The tissue elasticity and Yang's modulus value were recorded and compared.

Results:The tissue elasticity chart of patients in the treatment group before treatment was mainly greenish blue with the score of 3.70±1.53,and the Yang's modulus was(43.4±15.6) kPa. The tissue elasticity figure after treatment was mainly blue with the score of 2.40±0.87,and the Yang's modulus was(29.0±5.9) kPa. Whereas in the control group,the tissue elasticity figure was mainly blue with the score of 1.60±0.72,and the Yang's modulus was (24.0±7.6) kPa. These were statistical differences between the two groups(P=0.000).

Conclusion:SWE can be used as an evaluation method of manipulation treatment for neck-shoulder myofascial pain syndrome,which is an objective and sensitive detection method.
KEY WORDS  Neck  Shoulder  Myofascial pain syndromes  Manipulation,orthopedic  Shear wave elastography(SWE)
 
引用本文,请按以下格式著录参考文献:
中文格式:郭玲,张晨,张丁丁,高景华,刘广会,王尚全.剪切波超声弹性成像技术在颈肩肌筋膜疼痛综合征中的应用[J].中国骨伤,2016,29(2):142~145
英文格式:GUO Ling,ZHANG Chen,ZHANG Ding-ding,GAO Jing-hua,LIU Guang-hui,WANG Shang-quan.Application of shear wave elastography in the evaluation of neck-shoulder myofascial pain syndrome[J].zhongguo gu shang / China J Orthop Trauma ,2016,29(2):142~145
阅读全文  下载  查看/发表评论  下载PDF阅读器
关闭




版权所有:《中国骨伤》杂志社京ICP备12048066号-2  版权声明
地址:北京市东直门内南小街甲16号,100700
电话:010-64089487 传真:010-64089792 Email:zggszz@sina.com

京公网安备 11010102004237号