满足日常生活活动范围的髋臼假体角度安全范围及杯颈前倾角组合
摘要点击次数: 2577   全文下载次数: 1197   投稿时间:2011-04-20    
作者Author单位AddressE-Mail
李永奖 LI Yong-jiang 温州医学院附属第三医院骨科,浙江 瑞安 325200 Department of Orthopaedics, the Third Affiliated Hospital of Wenzhou Medical College, Ruian 325200, Zhejiang, China  
蔡春元 CAI Chun-yuan 温州医学院附属第三医院骨科,浙江 瑞安 325200 Department of Orthopaedics, the Third Affiliated Hospital of Wenzhou Medical College, Ruian 325200, Zhejiang, China  
张力成 ZHANG Li-cheng 温州医学院附属第三医院骨科,浙江 瑞安 325200 Department of Orthopaedics, the Third Affiliated Hospital of Wenzhou Medical College, Ruian 325200, Zhejiang, China jointlyj@126.com 
杨国敬 YANG Guo-jing 温州医学院附属第三医院骨科,浙江 瑞安 325200 Department of Orthopaedics, the Third Affiliated Hospital of Wenzhou Medical College, Ruian 325200, Zhejiang, China  
周德彪 ZHOU De-biao 温州医学院附属第三医院骨科,浙江 瑞安 325200 Department of Orthopaedics, the Third Affiliated Hospital of Wenzhou Medical College, Ruian 325200, Zhejiang, China  
彭茂秀 PENG Mao-xiu 温州医学院附属第三医院骨科,浙江 瑞安 325200 Department of Orthopaedics, the Third Affiliated Hospital of Wenzhou Medical College, Ruian 325200, Zhejiang, China  
陈文良 CHEN Wen-liang 温州医学院附属第三医院骨科,浙江 瑞安 325200 Department of Orthopaedics, the Third Affiliated Hospital of Wenzhou Medical College, Ruian 325200, Zhejiang, China  
赵章伟 ZHAO Liang-wei 温州医学院附属第三医院骨科,浙江 瑞安 325200 Department of Orthopaedics, the Third Affiliated Hospital of Wenzhou Medical College, Ruian 325200, Zhejiang, China  
期刊信息:《中国骨伤》2011年,第24卷,第11期,第930-934页
DOI:10.3969/j.issn.1003-0034.2011.11.014
基金项目:浙江省科技计划基金资助项目(编号:2009C33144);温州市科技计划基金资助项目(编号:Y20090274)
中文摘要:

目的:研究满足日常生活活动范围条件下,全髋关节置换术中髋臼杯角度安全范围及杯颈前倾角优化组合。

方法:建立人工全髋关节三维可视化计算机模型,将髋关节屈曲≥110°、屈曲90°内旋≥30°、后伸≥30°、外旋≥40°定义为日常生活活动范围一般标准,而将屈曲≥120°、屈曲90°内旋≥45°、后伸≥30°、外旋≥40°定义为严格标准,头颈直径比为2~2.92,假体颈前倾角为0°~30°,髋臼杯前倾角为0°~70°,外展角为10°~60°,颈干角设定为135°,计算满足上述两组活动范围,髋臼杯外展角每变化5°,相应的髋臼杯前倾角安装角度,在以臼杯外展角为横坐标,臼杯前倾角为纵坐标的坐标系上描点、连线画图进行描述,采用SAS 6.12统计学软件对数据进行分析。

结果:髋臼杯角度安全范围随着头颈直径比增大而增大,头颈比大于2.37时,臼杯角安全范围增大明显,严格标准活动度下的安全范围比一般标准条件下的小。颈干角135°,髋关节一般标准和严格标准活动度下,髋臼杯前倾角(Y)与股骨假体前倾角(X)的关系分别为:Y1=-0.816X1+39.76(R2=0.993),Y2=-0.873X2+47.04(R2=0.999).

结论:头颈直径比较大的假体髋臼杯角安全范围明显增大,建议选择头颈比直径大于2.37的假体。髋关节活动范围要求越高,髋臼杯角度安全范围就越小,但可以通过选择较大头颈直径比的假体来纠正。杯颈前倾角呈负相关。
【关键词】关节成形术,置换,髋  髋假体  活动范围,关节  计算机,模拟
 
The cup safe-zone and optimum combination of the acetabular and femoral anteversions that fulfills the desired range of motion of the hip
ABSTRACT  

Objective: To investigate safe range of acetabular prosthesis angle and the optimum combination of cup and neck in the range of activities of daily living(ADL).

Methods: A three-dimensional generic parametric and kinematic simulation module of THA was developed. Range of motion(ROM) of flexion ≥110°, internal-rotation ≥30° at 90° flexion, extension ≥30° and external rotation ≥40° were defined as the normal criteria for desired ROM for activities of daily living(ADL), and ROM of flexion ≥120°, internal-rotation ≥45° at 90° flexion, extension ≥30° and external rotation ≥40° as the severe criteria. The range of the changes in the general head-neck ratios(GR), the femoral neck anteversion (FA), the operative anteversion(OA) and operative inclination(OI) of the cup component were 2-2. 92, 0°-30°, 0°-70°, 10°-60°respectively. For the collodiaphyseal angle(CCD) of 135°, the corresponding OA related to the OI which every 5° was calculated, and described using dots and lines in a coordinate system in which OI of acetabular cup was the horizontal ordinate and OA of acetabular cup was the vertical ordinate. All data was analyzed by SAS 6. 12 software.

Results: Large GR greatly increased the size of safe range and it was recommended that GR be more than 2. 37 when the CCD-angle was 135° as it further increases the size of safe-zone. The size of cup safety range of the severe criteria was smaller than normal criteria. When the CCD-angle was 135°, the optimum relationship between OA of acetabular and FA of the normal criteria and the severe criteria can be estimated by the formula:Y1=-0. 816X1+39. 76(R2=0. 993), Y2=-0. 873X2+47. 04(R2=0. 999) respectively.

Conclusion: Large GR greatly increases the size of safe-zone and it is recommended that GR be more than 2. 37, so it extends the acceptable range of error that clinicians cannot avoid it completely. The larger range of the hip motion, the smaller size of the cup safe-zone, but can retrieve by increase the GR. The OA is negative with FA.
KEY WORDS  Arthroplasty, replacement, hip  Hip prosthesis  Range of motion, articular  Computers, analog
 
引用本文,请按以下格式著录参考文献:
中文格式:李永奖,蔡春元,张力成,杨国敬,周德彪,彭茂秀,陈文良,赵章伟.满足日常生活活动范围的髋臼假体角度安全范围及杯颈前倾角组合[J].中国骨伤,2011,24(11):930~934
英文格式:LI Yong-jiang,CAI Chun-yuan,ZHANG Li-cheng,YANG Guo-jing,ZHOU De-biao,PENG Mao-xiu,CHEN Wen-liang,ZHAO Liang-wei.The cup safe-zone and optimum combination of the acetabular and femoral anteversions that fulfills the desired range of motion of the hip[J].zhongguo gu shang / China J Orthop Trauma ,2011,24(11):930~934
阅读全文  下载  查看/发表评论  下载PDF阅读器
关闭




版权所有:《中国骨伤》杂志社京ICP备12048066号-2  版权声明
地址:北京市东直门内南小街甲16号,100700
电话:010-64089487 传真:010-64089792 Email:zggszz@sina.com

京公网安备 11010102004237号