前下后上型齿突骨折螺钉内固定的生物力学研究 |
摘要点击次数: 2496
全文下载次数: 1355
投稿时间:2007-12-17
|
作者 | Author | 单位 | Address | E-Mail |
方弘伟 |
FANG Hong-wei |
金华市中医医院骨科,浙江 金华 321017 |
Department of Orthopaedics,Traditional Chinese Medical Hospital of Jinhua,Jinhua 321017,Zhejiang,China |
HWF25888@163.com |
李建军 |
LI Jian-jun |
金华市中医医院骨科,浙江 金华 321017 |
Department of Orthopaedics,Traditional Chinese Medical Hospital of Jinhua,Jinhua 321017,Zhejiang,China |
|
吴增晖 |
WU Zeng-hui |
广州军区总医院 |
|
|
欧阳钧 |
OU-YANG Jun |
广州第一军医大学 |
|
|
钟世镇 |
ZHONG Shi-zhen |
广州第一军医大学 |
|
|
林文录 |
LIN Wen-lu |
金华市公安局 |
|
|
蒋良俊 |
JIANG Liang-jun |
金华市公安局 |
|
|
赵俊强 |
ZHAO Jun-qiang |
金华市中医医院骨科,浙江 金华 321017 |
Department of Orthopaedics,Traditional Chinese Medical Hospital of Jinhua,Jinhua 321017,Zhejiang,China |
|
于继华 |
YU Ji-hua |
金华市中医医院骨科,浙江 金华 321017 |
Department of Orthopaedics,Traditional Chinese Medical Hospital of Jinhua,Jinhua 321017,Zhejiang,China |
|
王向阳 |
WANG Xiang-yang |
温州医学院附属第二医院 |
|
|
|
期刊信息:《中国骨伤》2008年,第21卷,第4期,第279-281页 |
DOI:doi:10.3969/j.issn.1003-0034.yyyy.nn.zzz |
基金项目: |
|
中文摘要:
目的:评价前下后上型齿突骨折螺钉内固定的生物力学效果。
方法:收集20具新鲜成人上颈椎标本,分为3组,模拟齿突II型骨折模型,骨折线为前下后上,截骨角度分别为0° 6具,17° 8具,25° 6具,并进行齿突螺钉固定。进行旋入力矩、最大轴向拔出力和刚度测试。
结果:各组最大螺钉旋入力矩、抗拔出力差异无统计学意义。骨折端错位程度随着截骨角度的增加而增加,在17°组平均移位1.61 mm,25°组平均移位2.88 mm.抗扭转力矩随着截骨角度的增加而增加。各组剪切刚度差异无统计学意义。
结论:一定角度范围内的前下后上型骨折可使用齿突螺钉固定进行治疗。 |
【关键词】颈椎 齿突尖 骨折 骨折固定术,内 生物力学 |
|
A biomechanical study of anterior screw fixation for type II odontoid fracture with anteroinferior-posterosuperior fracture line |
|
ABSTRACT
Objective: To investigate the biomechanical effect of anterior screw fixation on the type II fractured odontoid process.
Methods: Twenty fresh human C1-C2 vertebrae specimens were harvested and randomly divided into three groups. The angle of type II fracture line was 0° in group I(n=6),17° in group II(n=8) and 25° in group III(n=6). The fractures were treated by anterior screw fixation. Insertion torque,maximal axial pullout force and stiffness of the bone-screw were tested.
Results: There was no significant difference of screw insertion torque and the pull-out strength between each group. The displacement of the odontoid fragment had an association to the angle of the fracture line,the displacement of the small angle was significantly higher than that of the large one(P<0.5). No significant difference of structure stiffness of the bone-screw was found between e- ach group.
Conclusion: Anterior screw fixation is feasible for type II odontoid fracture with certain fracture line extends from anteroinferior to posterosuperior. |
KEY WORDS Cervical vertebrae Odontoid process Fractures Fracture fixation,internal Biomechanics |
|
引用本文,请按以下格式著录参考文献: |
中文格式: | 方弘伟,李建军,吴增晖,欧阳钧,钟世镇,林文录,蒋良俊,赵俊强,于继华,王向阳.前下后上型齿突骨折螺钉内固定的生物力学研究[J].中国骨伤,2008,21(4):279~281 |
英文格式: | FANG Hong-wei,LI Jian-jun,WU Zeng-hui,OU-YANG Jun,ZHONG Shi-zhen,LIN Wen-lu,JIANG Liang-jun,ZHAO Jun-qiang,YU Ji-hua,WANG Xiang-yang.A biomechanical study of anterior screw fixation for type II odontoid fracture with anteroinferior-posterosuperior fracture line[J].zhongguo gu shang / China J Orthop Trauma ,2008,21(4):279~281 |
|
阅读全文 下载 查看/发表评论 下载PDF阅读器 |
关闭 |
|
|
|