影响骨外固定刚度和应力分布因素的实验研究
摘要点击次数: 1768   全文下载次数: 922   投稿时间:1995-10-06    
作者Author单位AddressE-Mail
赵建华 Zhao Jian-hua 第三军医大学大坪医院 The Third Affiliated Hospital,The Third Military Medical University,Chongqing 630042  
李起鸿 Li Qi-hong 第三军医大学西南医院 The Third Affiliated Hospital,The Third Military Medical University,Chongqing 630042  
期刊信息:《中国骨伤》1996年,第9卷,第6期,第9-11页
DOI:doi:10.3969/j.issn.1003-0034.yyyy.nn.zzz
基金项目:
中文摘要:本实验通过体外力学测试研究固定钢针数目、针直径和骨断端加压固定等因素对骨外固定刚度和应力分布的影响。实验结果表明:增多固定针数,增大针直径能显着提高骨外固定刚度(P<0.05),同时也非常显着地增大了应力遮挡率(P<0.001);骨断端加压固定既能显着提高骨外固定刚度(P<0.05),又能非常显着地降低应力遮挡率(P<0.0001).说明可以通过调整固定钢针数目、针直径和骨断端加压量,使骨外固定刚度和应力分布最适宜骨折愈合。
【关键词】骨外固定  生物力学
 
The exprimental research on the influential factors of rigidity and stress distribution at the external skeletal fixation
ABSTRACT  The influence of the number and diameter of the stainless steel pin and the compression at the external skeletal fixation were studied with external mechonical examination.The conclusions are as follows;(1) increasing the number and diameter of pin can not only improve the stiffness of external skeletal fixation, but also significantly increase the rate of stress-protection;(2) compression of the external skeletal fixation can not only improve the stiffness of the external skeletal fixation,but also reduce the rate of stress-protection significantly. It proved that the aims of immobilization of the bone fragments and promotion of fractureunion can be achieved by adjusting the number and diameter of pin and the compression at fracture ends.
KEY WORDS  Exetrnal skeletal fixation  Biomechnics
 
引用本文,请按以下格式著录参考文献:
中文格式:赵建华,李起鸿.影响骨外固定刚度和应力分布因素的实验研究[J].中国骨伤,1996,9(6):9~11
英文格式:Zhao Jian-hua,Li Qi-hong.The exprimental research on the influential factors of rigidity and stress distribution at the external skeletal fixation[J].zhongguo gu shang / China J Orthop Trauma ,1996,9(6):9~11
阅读全文  下载  查看/发表评论  下载PDF阅读器
关闭




版权所有:《中国骨伤》杂志社京ICP备12048066号-2  版权声明
地址:北京市东直门内南小街甲16号,100700
电话:010-64089487 传真:010-64089792 Email:zggszz@sina.com

京公网安备 11010102004237号